For performance of ISBT Method 10.0 THC content of bev-grade CO2 and other gas applications. The Model 9000 Total Hydrocarbon Analyzer (FID) System with Automatic Calibration Control.
- 1 Point (Span/Zero/Sample)
- 5 programmable relay outputs
- 4-20 mA analog outlet
- RS-232
- Ethernet digital outputs
- 9000 Keeper Software Package
- Methods Manuals included
Insturment Requirements
Power:
- Requires 90 - 230VAC, 50-60 Hz.
Fuel Gases:
- Hydrogen fuel gas (Ultra Pure)
- 30 psig delivery
- 35 cc/min (Electronic Flow Controlled)
- zero-grade air flow
- 30 psig
- 175 cc/min flow (Electronic Flow Controlled)
Basic CO2 software method is pre-programmed by Airborne Labs International as default. Certified laboratory back-up testing also available for result confirmation.
Detector: |
Flame Ionization Detector (FID) |
Ranges: |
User definable based upon calibration within: MDQ 0.01ppm, full-scale 1-200 ppm (Methane) MDQ 0.1ppm, full-scale 1-2,000ppm (Methane) MDQ 0.3ppm, full-scale 1-20,000ppm (Methane) MDQ 0.003%, full-scale 0.01-100% (Methane) |
Repeatability: |
± 1% full-scale response |
Drift: |
± 1% of full-scale over 24 hours |
Response Time: |
< 5 seconds to 90% of final reading |
Alarms: |
Multilevel concentration and fault alarms that result in audible and visual alarms. Alarms may also be mapped to relays to control external equipment |
Sampling: |
Internal single or multipoint modules, for pre-filtered (<0.1 microns) non-condensing samples, with or without sample pump |
Optional Accessories:
- Hydrogen Generator
- Zero Air Generator
- Digitial Flow Meter
- Hydrogen Leak Detector
- Certified Gas Standards
- Pressure Regulators and transfer lines
|