The BAM-8900 is a SilcoNert® passivated Benzene (Bz), Acetaldehyde (AA) and optional Methanol (MeOH) Analyzer.
8900 BAM |
|
Analysis Time: |
< 360 seconds |
Detector: |
PID (High-sensitivity PID Optional) |
Column: |
Capillary |
Oven Temperature: |
100°C, Nominal |
Carrier Gas: |
Hydrogen, 15 cc/min |
Lower Detection Limit: |
< 50 ppb Acetaldehyde, < 2 ppm Methanol, < 2 ppb Benzene |
Accuracy: |
1% of full-scale |
Precision: |
2% of measured value |
Span Drift (25 hrs.): |
< 2% of full-scale |
Sample Flow Rate: |
250 - 500 cc/min, typical |
Analog Outputs: |
(1) 0 - 20 mA or 4 - 20 mA loop power supplied, isolated. Selectable for: gas concentration, unintegrated detector signal. Options for up to 20 additional programmable 0 - 20 mA, 4 - 20 mA or voltage outputs: 0 - 1V, 0 - 5V, or 0 - 10V |
Digital Outputs: |
RS-232, optional Local Area Network |
Relays: |
(5) User programmale relays for concentration and diagnostic alarms (1A @ 30Vdc). Options for up to 32 additional relays available. |
Inputs: |
Optional digital input board for 3 contact closure inputs. Supports start analysis, start calibration, and analyze calibration gas sample. |
Fuel Gases:
- Hydrogen Carrier gas (Ultra Pure)
- Actuator Air
Basic CO2 software method is pre-programmed by Airborne Labs International as default. Certified laboratory back-up testing also available for result confirmation.
Optional Accessories:
- Hydrogen Generator
- Zero Air Generator
- Digitial Flow Meter
- Hydrogen Leak Detector
- Certified Gas Standards
- Pressure Regulators and transfer lines
|