Ranges: |
0 - 50 ppb to 0 - 20,000 ppb full-scale, user selectable; dual ranges and auto-ranging supported |
Zero Noise: |
< 0.2 ppb |
Span Noise: |
< 0.5% of reading (above 50 ppb) |
LDL: |
0.4 ppb |
Zero Drift: |
<0.5 reading / 24 hours; 1 ppb / 7 days |
Span Drift: |
<0.5% reading / 24 hours; 1% of reading / 7 days |
Lag Time: |
20 seconds |
Rise Time (95%): |
< 120 seconds |
Linearity: |
1% of full scale |
Sample Flow Rate: |
700 cc/min ± 10% |
Converter Temperature: |
1000°C |
Converter Efficiency: |
> 98% above 0°C dewpoint |
Converter Life: |
3000 ppm hours |
SOx Scrubber Efficiency: |
> 98% |
SOx Scrubber Life: |
> 1000 ppm hours |
Operating Temperature: |
5 to 40°C |
Power: |
100 VAC or 240 VAC, 50/60 Hz (user specified) |
Outputs: |
Analog: 10V, 5V, 1V, 100mV (selectable)
(4 - 20 mA isolated output optional)
RS232 (I\O): Standard DB9
Digital Status: 12 contact closure outputs |
Size (H x W x D): |
Model 6200E analyzer:
(7" x 17" x 23.5") (177.8 x 432 x 596.9 mm)
Dilutor / Converter Module:
(7" x 17" x 23.5") (177.8 x 432 x 596.9 mm)
Calibration Module:
(5.25" x 17" x 23.5") (133.3 x 432 x 596.9 mm)
|